MAUDE Adverse Event Report: STAAR SURGICAL EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS; PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS; PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_12.6 (-10.0/2.0)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/03/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced a low vault with rotation.In a separate visit the lens was exchanged for a longer length lens.The replacement lens resolved the problem.Cause reported as a patient related factor.
 
Additional Manufacturer Narrative
Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key24346718
Report Number2023826-2026-00242
Device Sequence Number10041066
Product Code QCB
UDI-Device Identifier00840311308385
UDI-Public(01)00840311308385(11)250814(17)280731
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVTICM5_12.6 (-10.0/2.0)
Device Catalogue NumberVTICM5_12.6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2026
Initial Report FDA Received Date02/16/2026
Date Device Manufactured08/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45: LOT# UNK; FOAM TIP PLUNGER (FTP), LOT# UNK; INJECTOR MODEL: MSI-PF, LOT# UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 YR
Patient SexFemale
-
-