BOSTON SCIENTIFIC CORPORATION COYOTE? ES; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number H74939135202010
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/23/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the moderately tortuous and moderately calcified superficial femoral artery.A 2mm x 20mm x 143cm coyote? es balloon catheter was advanced for dilatation.During the procedure, the balloon ruptured after the second inflation at 14 atmospheres for 5 seconds.The device was removed without any problem, and the procedure was completed with a different device.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE? ES
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key24349879
Report Number2124215-2026-08659
Device Sequence Number11951851
Product Code LIT
UDI-Device Identifier08714729767176
UDI-Public(01)08714729767176(17)280709(10)36941709
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K093636
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH74939135202010
Device Catalogue NumberH74939135202010
Device Lot Number0036941709
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2026
Initial Report FDA Received Date02/16/2026
Date Device Manufactured07/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-