| Model Number |
VICM5 13.2 (-5.00) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
11/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: one similar complaint type events reported for units within the same lot.Manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced refractive surprise.The lens was exchanged with a same length but different power lens.The cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H6- 10: returned product evaluation - the subject lens was received dry within a microcentrifuge tube.Visual inspection found an optic and haptic bent lens.Dimensional and functional inspection found the lens to be within specifications.Claim#(b)(4).
|
| |
|