| Model Number |
SIGADAPTXL |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907); Difficult to Open or Close (2921); Human-Device Interface Problem (2949); Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: sigphandle, sig power sigphandle handle, (serial number: (b)(6) egia45amt, egia45amt egia 45 artic med thick sulu, (lot number: unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during an operation, the adapter could be fired with two reloads; however, with the third magazine, the reload jaws would not open and close.Instead, the magazine angled and straightened.Attempts to change the reload were unsuccessful, as the reload could not be disconnected from the adapter and ultimately broke off.The operation was continued using a new adapter and a new magazine.The same handle worked fine.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.It was reported that the reload could not be disconnected from the adapter.The reported issue was confirmed.The most likely cause was traced to non-device related factors.It was also reported that the jaws of the third reload would not open; instead the jaws angled and straightened.The reported issues could not be confirmed.The most likely cause could not be established from the information available.It was additionally reported that the reload broke off when being unloaded.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|