• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 1829500
Medical Device Problem Code Expulsion (2933)
Health Effect - Clinical Codes Skin Erosion (2075); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/26/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
A patient underwent a port placement procedure using the powerport m.R.I.Duo.Post procedure, the patient allegedly experienced soreness on the right side of the port for one month.The port reportedly protruded due to the patient¿s thin body.Additionally, the patient experienced a sensation that her heart felt different, and at one point her heart rate allegedly dropped to 43 for a second.The current status of the patient is unknown.
 
Event or Problem Description
A patient underwent a port placement procedure using the powerport m.R.I.Duo.Post procedure, the patient allegedly experienced soreness on the right side of the port for one month.The port reportedly protruded due to the patient¿s thin body.Additionally, the patient experienced a sensation that her heart felt different, and at one point her heart rate allegedly dropped to 43 for a second.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.The investigation is inconclusive for the reported expulsion as no objective evidence was provided for review.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D4 (unique identifier (udi) #), g3, h6 (patient, component, method).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key24361888
Report Number3006260740-2026-00878
Device Sequence Number18671616
Product Code LJT
UDI-Device Identifier00801741027185
UDI-Public(01)00801741027185(11)220922(17)231031(10)REGT5298
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K090512
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number1829500
Device Lot NumberREGT5298
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2026
Supplement Date Received by Manufacturer04/09/2026
Initial Report FDA Received Date02/17/2026
Supplement Report FDA Received Date04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-