| Catalog Number |
451570107 |
| Medical Device Problem Codes |
Device Slipped (1584); Device-Device Incompatibility (2919); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Event or Problem Description
|
|
It was reported that during set-up for a total knee arthroplasty (tka) surgical procedure it was observed that 2x robotic assisted array clamp devices were coming loose on their own.There were no adverse consequences to the patient.No additional information could be provided.This is report 1 of 2 for (b)(4).
|
| |
|
Event or Problem Description
|
|
Updated event description: it was reported that during set-up for a total knee arthroplasty (tka) surgical procedure it was observed that 3x robotic assisted array clamp devices were coming loose on their own.There were no adverse consequences to the patient.No additional information could be provided.This is report 1 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: correction b5: additional event information was received.The reporter stated that 3 devices were involved in the event.Please note that the event description updated.D4: the lot number and expiration date were reported as unknown on the initial report; and have been updated accordingly.Therefore, the udi has been updated accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: h6: please note that the medical device problem code has been updated.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: investigation summary: the product was returned to depuy synthes for evaluation.Visual analysis of the returned device revealed that the velys array clamp presents an overall worn appearance consistent with multiple uses.No signs of damage nor defects that could have affect the functionality were observed.None of the threads that were visible appeared damaged.Functional testing was conducted using the returned device set and a product with the equivalent diameter of a velys fixation pin.The wedge component of the array clan bone assembly was able to be fully tightened through the wedge wingnut.The device was able to retain the pin as intended.The wing knob component that connects the two pieces together was able to be fully tightened and the teeth meshed as intended.The wing knob component that tightens onto the pins can be fully tightened both with and without pins inserted.The array clamp array clasp assembly button was fully functional in securing, retaining, and releasing the production equivalent velys arrays without difficulty.The overall complaint was not confirmed as the observed condition of the velys array clamp would not have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Search Alerts/Recalls
|