• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ATOS MEDICAL AB PROVOX ACTIVALVE; TRACHEOESOPHAGEAL SPEECH VALVE, INDWELLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATOS MEDICAL AB PROVOX ACTIVALVE; TRACHEOESOPHAGEAL SPEECH VALVE, INDWELLING Back to Search Results
Catalog Number 7161
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Aspiration/Inhalation (1725); Localized Skin Lesion (4542)
Date of Event 12/18/2025
Type of Reportable Event Serious Injury
Event or Problem Description
According to the available information, this complaint concerns an end user of a provox activalve strong voice prosthesis (22.5fr/6mm) during use the device was placed in (b)(6) 2025.Approximately four months later, the patient reported to the speech language pathologist (slp) that the small blue center piece detached while drinking and was swallowed by the patient.The patient reported feeling the piece, spitting it back up, and then experiencing choking during the next sip of liquid.The patient reported that without the small blue center piece, he was choking and unable to voice/communicate.The speech language pathologist (slp) evaluated the patient the following morning in clinic, removed the device, and placed a new, different device.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROVOX ACTIVALVE
Common Device Name
TRACHEOESOPHAGEAL SPEECH VALVE, INDWELLING
Manufacturer (Section D)
ATOS MEDICAL AB
1 holtedam
humlebæk 3050
DA  3050
Manufacturer (Section G)
HÖRBY
Manufacturer Contact
usgab giovanna hayford-quansah
1601 west river road n
minneapolis, MN 55411
MDR Report Key24365489
Report Number3002806269-2026-00016
Device Sequence Number14756075
Product Code EWL
UDI-Device Identifier07331791000584
UDI-Public7331791000584
Combination Product (Y/N)N
PMA/510(K) Number
K092593
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number7161
Device Lot Number2501098_L716101000
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/13/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Other;
Patient SexUnknown
-
-