• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA PLATELET+SAMPLER, PLASMA, RBC SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA PLATELET+SAMPLER, PLASMA, RBC SET Back to Search Results
Catalog Number 4824462
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation: the run data file (rdf) was analyzed for this event. the signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was possibly the result of a wbc saturation of the lrs chamber that fell just under the detection threshold for flagging the platelet product for wbc verification, which occurred just prior to the scheduled chamber clearing.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product may have been donor-related investigation is in process, a follow-up report will be provided.
 
Event or Problem Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.Donor unit #: (b)(6) there was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.11.Investigation: the run data file (rdf) was analyzed for this event.The signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was possibly the result of a wbc saturation of the lrs chamber that fell just under the detection threshold for flagging the platelet product for wbc verification, which occurred just prior to the scheduled chamber clearing.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product may have been donor-related the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Root cause: the signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was possibly the result of a wbc saturation of the lrs chamber that fell just under the detection threshold for flagging the platelet product for wbc verification, which occurred just prior to the scheduled chamber clearing.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product may have been donor-related.
 
Event or Problem Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.Donor unit #: (b)(6) there was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Common Device Name
TRIMA PLATELET+SAMPLER, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key24367410
Report Number1722028-2026-00032
Device Sequence Number17488095
Product Code GKT
UDI-Device Identifier05020583824461
UDI-Public05020583824461
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
BK190332
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 02/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4824462
Device Lot Number2509291242
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2026
Supplement Date Received by Manufacturer03/27/2026
Initial Report FDA Received Date02/17/2026
Supplement Report FDA Received Date04/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-