| Catalog Number |
82005 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
In this event it is reported that a 30k fsi-sli-fg-10s ins, pkd alleged has no water flow.No injury occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results multiple unsuccessful attempts were made to obtain the device for evaluation.Dhr documentation has been reviewed.There were no deviations identified during the manufacturing process of this lot that could have resulted in this alleged event.Failure mode - water flow issue root cause - no defect conclusion code - no failure found.
|
| |
|