|
The reported complaint that a balloon leakage was suspected on the solex catheter (lot 207977) was not confirmed during the visual and functional testing.No issues or discrepancies were found.No leak, no damage, no device malfunction was observed during testing, and the catheter functioned as intended.Visual inspection of the returned catheter was performed.As returned, the guidewire was inside the catheter, and the guidewire was able to be removed from the catheter without resistance.There was no physical damage observed on the catheter.Blood residue was observed on the catheter's serpentine balloon.Functional testing of the returned catheter was performed.All infusion ports and lumens were flushed without resistance.During functional pressure leak test, the catheter was connected to a pressurized inflation device, and the balloon was fully inflated when pressurized up to 100 psi; no leak, no issues were found on the catheter.The balloon did not leak during testing.In addition, the returned catheter was connected to a known-good suk and ran on a peristaltic pump for 10 minutes at a flow rate of 240 ml/min.No leak was found, and the catheter functioned as intended.During further functional testing, the returned catheter was connected to a known-good suk and ran on the thermogard console system, running in the max warming mode with a target temperature at 37°c for 60 minutes and running in the max cooling mode with a target temperature at 35°c for 60 minutes.No leak was observed on the catheter.The balloon was properly warming during the warming mode and cooling during the cooling mode.The red pinwheel of the suk was spinning as normal.No problem was found, and the catheter functioned as intended.
|