• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH BUTTRESS/COMPR-NUT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH BUTTRESS/COMPR-NUT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 03.037.016
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported that on 02 february, 2026 upon disengaging tfna lag screw drill/insert sleeve, the outer sleeve and buttress wheel would not disengage.Surgeon and scrub tech used a mallet to release it and caused the buttress wheel to cross thread with the outer sleeve.Surgery was delayed for 5 minutes due to the reported event.Malleted sleeve and wheel out, procedure was successfully completed.No fragments were generated.No patient consequences.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d4 (lot, expiration date), h4 corrected: d4 primary udi number if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: corrected: g1, h4.Added: d10.H3, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found nut is stripped at the distal end.An interaction was made between the buttress/compr-nut and the blade/screw guide sleeve and was found that devices were unable to be disassembled given the condition of the blade/screw guide sleeve, therefore, the reported condition was able to be replicated.The observed condition of the device is consistent with a component failure that was caused by exposure to unintended forces like overtightening the device while assemblance; or by assembling the device in a misaligned position.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the buttress/compr-nut would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review.Part number: 03.037.016-us.Lot number: 96p4881.Manufacturing site: hägendorf.Release to warehouse date: 28-april-2021.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BUTTRESS/COMPR-NUT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
WERK HAEGENDORF
im bifang 6
haegendorf CO
SZ  
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key24369908
Report Number8030965-2026-01572
Device Sequence Number18587441
Product Code LXH
UDI-Device Identifier10886982070326
UDI-Public(01)10886982070326(17)411231(10)96P4881
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.037.016
Device Lot Number96P4881
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/02/2026
Supplement Date Received by Manufacturer02/18/2026
03/18/2026
Initial Report FDA Received Date02/17/2026
Supplement Report FDA Received Date02/23/2026
03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BLADE/SCREW GUIDE SLEEVE
Patient SexUnknown
-
-