| Catalog Number |
51-106150 |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during inspection of circulating stock, a sterile package was found damaged.There was no patient involvement.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign, event occurred in japan.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: d9, g3; h2; h3; h6, h10.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (pouch).Sterility has not been compromised.Therefore, this is no longer considered reportable.Root cause has been identified as transit damage and a packaging design issue.Actions have been previously raised to implement improved packaging changes.These packaging changes will reduce the number of transit related packaging damage events.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|