|
It was reported that a patient with an inflatable penile prosthesis (ipp) presented with asymmetry, as one side appeared longer than the other, and the device did not seem to function correctly.A revision procedure was performed, during which the physician suspected a possible proximal erosion on the left side and a crossover; however, no erosion was confirmed.During surgery, the crossed over cylinder was explanted, the affected corporal space was remeasured, and it was determined that a longer rear tip extender was required.A 2 cm rte was replaced with a 4cm rte to match the right side.No patient complications were reported.
|
|
Udi field (d4) concomitant product udi for reservoir is (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.Migration, length too short, mechanical issue and tissue damage are acknowledged within the dfu and are not related to off-label use or failure to follow instructions.The reported patient symptom of tissue damage is a known risk associated with this device type and indicated as such in the instructions for use.
|