MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE; SMALL INTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE; SMALL INTESTINAL VIDEOSCOPE Back to Search Results
Model Number SIF-Q180
Medical Device Problem Code Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The legal manufacture reviewed the customer provided the cleaning disinfection and sterilization (cds) processes where no obvious deviations from instructions for use (ifu) were identified.The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the most probable cause of the malfunction could not establish.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the small intestinal videoscope had debris inside the lens.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE
Common Device Name
SMALL INTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24374280
Report Number9610595-2026-15582
Device Sequence Number4837917
Product Code FDA
UDI-Device Identifier04953170339943
UDI-Public04953170339943
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K071254
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberSIF-Q180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2026
Initial Report FDA Received Date02/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/05/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-