| Model Number |
SIGTRSB45AXT |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, during firing, the powered stapler would not clamp.Another reload was used with the same handle, shell, and adapter to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: sigphandle, sig power sigphandle handle, (serial#: (b)(6); sigpshell, sig power sigpshell control shell, (lot#: unknown); sigadaptstnd, sig power sigadaptstnd linear adapter, (serial#: (b)(6).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the jaw of the reload was open.It was reported that the jaws did not close.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: of device pre fired.The product analysis noted evidence that the device was not used as intended.This issue can occur if the firing button had been partially compressed and then the open button was pressed after pressing the green firing button.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to completely fire the reload will result in an incomplete cut and/or incomplete staple formation, which may result in poor hemostasis and/or leakage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|