• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN ENDO DISSECT; MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN ENDO DISSECT; MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED Back to Search Results
Model Number 176645
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, prior to use, the jaws were deformed and did not fit together.The issue occurred on seven devices.There was no patient involved.
 
Additional Manufacturer Narrative
D10 concomitant products: 176645 176645 endo dissect 254 - (lot#p5h0868); 176645 176645 endo dissect 254 - (lot#p5h0868); 176645 176645 endo dissect 254 - (lot#p5h0868); 176645 176645 endo dissect 254 - (lot#p5h0868); 176645 176645 endo dissect 254 - (lot#p5h0868); 176645 176645 endo dissect 254 - (lot#p5h0868); medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
D10 concomitant product: 176645, 176645 endo dissect 254 (lot#p5k1025); 176645, 176645 endo dissect 254 (lot#unknown) additional information: d9, g3, h3, h6 correction: a1 h3 evaluation summary: medtronic conducted an investigation based upon all information received.One representative sample was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the jaws misaligned.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO DISSECT
Common Device Name
MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED
Manufacturer (Section D)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24375405
Report Number2647580-2026-00422
Device Sequence Number5583807
Product Code GET
UDI-Device Identifier10884523000740
UDI-Public(01)10884523000740(17)300930(10)P5K1026
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K904578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number176645
Device Catalogue Number176645
Device Lot NumberP5K1026
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2026
Supplement Date Received by Manufacturer05/07/2026
Initial Report FDA Received Date02/18/2026
Supplement Report FDA Received Date05/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE NOTE ON H11; SEE H11
Patient SexUnknown
-
-