| Model Number |
176645 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, prior to use, the jaws were deformed and did not fit together.The issue occurred on seven devices.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products:(b))6) 176645 endo dissect 254 - (lot#p5h0868); (b)(6) 176645 endo dissect 254 - (lot#p5h0868); (b)(6) 176645 endo dissect 254 - (lot#p5h0868); (b)(6) 176645 endo dissect 254 - (lot#p5h0868); (b)(6) 176645 endo dissect 254 - (lot#p5h0868); (b)(6) 176645 endo dissect 254 - (lot#p5h0868); medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 correction: a1 h3 evaluation summary: medtronic conducted an investigation based upon all information received.One representative sample was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the jaws misaligned.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: 176645, 176645 endo dissect 254 (lot#p5k1025); 176645, 176645 endo dissect 254 (lot#unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|