• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX200T11
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
A device was returned to third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the third-party service center, the device was visually inspected and found evidence of foam degradation.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the third-party service center.There were no error codes found.The third-party service center concludes that they confirm the customer's allegation and there was a visible foam degradation and unit got scrapped.
 
Additional Manufacturer Narrative
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION CPAP
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key24375920
Report Number2518422-2026-102224
Device Sequence Number5572668
Product Code BZD
UDI-Device Identifier00606959424081
UDI-Public(01)00606959424081
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation 505
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 02/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberDSX200T11
Device Catalogue NumberDSX200T11
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2022
Initial Report FDA Received Date02/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/07/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1