|
While the customer was using a part of an intraoral sensor system, schick33 s2 sensor ship kit/3.0 cable 9', they allege experiencing image quality issues when an x-ray image was taken and an additional image was required.No injury was reported from the alleged event.
|
|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- troubleshooting description: tv connection, asked about tube head settings, 60kv\6ma.10, belmont photech 2, reviewed the technique factor chart.It was determined that the exposure level was adjusted and working better.
|