| Catalog Number |
IAB-06830-U |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn (b)(4).
|
| |
|
Event or Problem Description
|
|
It wsas reported "after the catheter inserted into the patient, the pump alarmed helium leakage 2.Change the new catheter".No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Event or Problem Description
|
|
It wsas reported "after the catheter inserted into the patient, the pump alarmed helium leakage 2.Change the new catheter".No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned for investigation was a was a 30cc 7.5fr ultraflex intra-aortic balloon catheter (iabc) without the original packaging carton.The sample was returned in a cardboard box and was in a bio-hazard bag.Upon return, the teflon sheath was noted on the returned iabc.The short arteri-al pressure tubing was noted connected to the iabc luer.The one-way valve was tethered to the short driveline tubing.The bladder was fully unwrapped.The central lumen was noted bent at approximately 5.1cm, 22cm and 66.8cm from the distal tip.The central lumen was noted broken at approximately 1.6cm from the iab distal tip.No blood was noted on the exterior surfaces of the returned sample.No obvious blood was noted within the helium pathway; the catheter was likely cleaned prior to return.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute according to quality system document.An attempt to aspirate and flush the iabc central lumen using a 60cc lab-inventory syringe was un-able to be successfully completed.Upon aspiration, water was unable to be consistently pulled into the syringe.The attempt to flush resulted in bladder inflation, which is consistent with the previously noted broken central lumen.The iabc was leak tested in accordance with testing methods from manufacturing procedure.A leak was immediately detected from the iabc distal tip and iabc luer end.The leak from the iabc distal tip and iabc luer end are consistent with the previously confirmed broken central lumen.The iabc was leak tested again with the iabc distal tip and luer end blocked off; no other leaks were detected.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.No resistance was noted; the guidewire exited the cen-tral lumen and entered the bladder at the location of the broken central lumen.Some blood was noted.The guidewire was front loaded through the iabc luer.The guidewire met resistance and could not advance at approximately 71.4cm from the luer; which is the location of the previously noted bend.Some blood was noted on the guidewire.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint for "pump alarmed helium leakage 2" is confirmed.During the investigation, the intra aortic balloon catheter (iabc) central lumen was noted broken near the iabc distal tip, which can cause a helium loss alarm.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the broken central lumen.The root cause of the complaint is unde-termined.No further action required at this time.This will be monitored for any developing trends.
|
| |
|
Search Alerts/Recalls
|