Additional information: d9, h3 plant investigation: a dialyzer from the reported lot was returned for evaluation.During gross visual examination, a potted fiber fragment was observed at approximately 120°, with the dialysate ports at 0°, on the non-cavity id end.Under magnification of 20x, the fiber fragment measured approximately 0.35mm in length.An opposing end was not identified.There was no other damage or irregularities noted on the returned sample.The reported issue is confirmed.The probable causes for this failure to occur are related to the handling of the fiber bundle during production and during the insertion process of the fiber bundle into the dialyzer housing.A review of the production record was performed.One non-conformance was opened during the production of this lot for exceeding the bubble point rejection rate.All discrepant product was discarded.There was no other indication of product non-acceptance or deviation in the manufacturing process related to the complaint event.This includes non-conformances, rework, labeling, process controls, and any other occurrence in production that was potentially related to the complaint.The reported lot number passed pyrogen testing, was within sterilization dosage parameters and passed all release criteria.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
|