MAUDE Adverse Event Report: OGDEN MANUFACTURING PLANT OPTIFLUX F250NRE HF 12/CS 2.5SA EBEAM; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OGDEN MANUFACTURING PLANT OPTIFLUX F250NRE HF 12/CS 2.5SA EBEAM; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 0500325E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 02/11/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A user facility inventory manager reported that a dialyzer blood leak occurred during the patient¿s hemodialysis (hd) treatment.Additional information was obtained during follow-up with a facility registered nurse (rn).The reported issue occurred immediately after the initiation of treatment.The blood leak was noted as being an internal blood leak.The leak was visually observed within the dialyzer.The fresenius 2008k2 machine alarmed appropriately with a blood leak alarm.Blood leak test strips were not used to test for the presence of blood.There was no defect or damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 250 ml.There was no patient injury or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be available to be returned to the manufacturer for evaluation.
 
Event or Problem Description
A user facility inventory manager reported that a dialyzer blood leak occurred during the patient¿s hemodialysis (hd) treatment.Additional information was obtained during follow-up with a facility registered nurse (rn).The reported issue occurred immediately after the initiation of treatment.The blood leak was noted as being an internal blood leak.The leak was visually observed within the dialyzer.The fresenius 2008k2 machine alarmed appropriately with a blood leak alarm.Blood leak test strips were not used to test for the presence of blood.There was no defect or damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 250 ml.There was no patient injury or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be available to be returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
Additional information: d9, h3 plant investigation: a dialyzer from the reported lot was returned for evaluation.During gross visual examination, a potted fiber fragment was observed at approximately 120°, with the dialysate ports at 0°, on the non-cavity id end.Under magnification of 20x, the fiber fragment measured approximately 0.35mm in length.An opposing end was not identified.There was no other damage or irregularities noted on the returned sample.The reported issue is confirmed.The probable causes for this failure to occur are related to the handling of the fiber bundle during production and during the insertion process of the fiber bundle into the dialyzer housing.A review of the production record was performed.One non-conformance was opened during the production of this lot for exceeding the bubble point rejection rate.All discrepant product was discarded.There was no other indication of product non-acceptance or deviation in the manufacturing process related to the complaint event.This includes non-conformances, rework, labeling, process controls, and any other occurrence in production that was potentially related to the complaint.The reported lot number passed pyrogen testing, was within sterilization dosage parameters and passed all release criteria.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX F250NRE HF 12/CS 2.5SA EBEAM
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key24377226
Report Number0001713747-2026-00085
Device Sequence Number5650481
Product Code KDI
UDI-Device Identifier00840861100170
UDI-Public00840861100170
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K162488
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0500325E
Device Lot Number25LU04024
Was Device Available for Evaluation? Device Returned to Manufacturer
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2026
Supplement Date Received by Manufacturer04/20/2026
05/28/2026
Initial Report FDA Received Date02/18/2026
Supplement Report FDA Received Date04/22/2026
05/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS 2008K2 MACHINE; FRESENIUS 2008K2 MACHINE; FRESENIUS BLOODLINES; FRESENIUS BLOODLINES
Patient Age41 YR
Patient SexMale
Patient Weight119 KG
Patient RaceWhite
-
-