This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 10, 4109, 3331 and 4111.H6: investigation findings code was added: 180.H6: investigation conclusions code was added: 4307.H11: narrative/data was updated.Zimvie received one the reported device for evaluation.Visual evaluation was performed.The implant has signs of use.There was blood and bone debris inside the implant internal threads.The implants drive feature was rounded and damaged.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and one other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was possibly user caused.Therefore, based on the available information, a device malfunction has occurred.The implants drive feature was rounded and damaged.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|