Model Number H74939171120470
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D2b - pro code (product code): fge, lit.G4 - premarket / 510(k) #: k141521, k141597.
 
Event or Problem Description
It was reported that a balloon rupture occurred.The 99% stenosed target lesion was located in the moderately tortuous and non-calcified left subclavian vein.A 12.0 x 40, 75cm mustang balloon catheter was advanced for dilation.During the procedure, the balloon ruptured upon first inflation at 8 atmospheres for 5 seconds.The device was removed without any problem, and the procedure was completed with another of the same device.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG?
Common Device Name
STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key24377841
Report Number2124215-2026-09060
Device Sequence Number11980713
Product Code FGE
UDI-Device Identifier08714729793441
UDI-Public(01)08714729793441(17)280720(10)36997789
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH74939171120470
Device Catalogue NumberH74939171120470
Device Lot Number0036997789
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2026
Initial Report FDA Received Date02/18/2026
Date Device Manufactured07/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE - RADIFOCUS; INFLATION DEVICE - NIPRO; INTRODUCER SHEATH - TERUMO 7F
Patient SexUnknown
-
-