| Catalog Number |
TSV4B13 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Abscess (1690); Unspecified Infection (1930); Osteolysis (2377)
|
| Date of Event |
12/05/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A2: age and date of birth unknown / not provided.A4: weight unknown / not provided.G4: additional pma/510(k) number ¿ k011028|k013227.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented due to pain on the implant bridge at tooth sites #21-23.It was noted that the patient had infection and mobility around the implants.The bridge was removed and the area was degranulated.The areas were allowed to heal and the patient opted to have a partial crafted, rather than having the implant bridge remade.Symptoms as a result of the event: abscess.
|
| |
|
Event or Problem Description
|
|
No additional event information received at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay corrected information, additional information and device evaluation.Correction: d6a: device implant date was submitted incorrectly as (b)(6) 2025.The correct device implant date is unknown.The following sections are being reported: b4: date of this report was updated.D6a device implant date was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 4110 h10: narrative/data was updated.Dhr, sterilization, and complaint history review could not be performed, as the subject lot number associated with the tsv4b13 is not available.Zimvie quality management system (qms) has controls in place to prevent the distribution of non-conforming product and ensure the product is within specifications.A complaint history review by item number was conducted for the tsv4b13 dating back to 12 months from now.The complaint history review revealed that there are no existing non-conformances/capa/hhe/d/product holds for the reported product for similar event.As documented in the summary investigations, contributing factors for the reported event likely exist outside of zimvie control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported events remain non-verifiable as they are a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for infection, bone loss, and/or peri-implantitis problems reported in association with implant failure have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|
| |
|
Search Alerts/Recalls
|
|
|