THERAKOS DEVELOPMENT LIMITED CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Catalog Number CLXUSA
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction centrifuge bowl leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot p308 was conducted.There were no non-conformances related to the complaint.This lot met all release requirements.A review of kit lot p308 shows no trends.Trends were reviewed for complaint category, centrifuge bowl leak/break.No trends were detected for this complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.A therakos field service engineer was dispatched to perform instrument service as a result of the complaint.A review of the service details verified the reported complaint; however, no contributing factors for the reported complaint were identified.No further action is required at this time.This investigation is now complete.Pqc-001513.N.S.(b)(6) 2026.
 
Event or Problem Description
The customer contacted therakos to report they experienced a centrifuge bowl leak/break with their cellex photopheresis kit("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported that the centrifuge bowl broke during the treatment after 1500mls of whole blood was processed.The ecp treatment was aborted and no residual blood within the kit was returned to the patient.The customer reported the patient was in stable condition.The kit return was requested; however, the customer discarded the kit.The customer did not return product for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Common Device Name
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS DEVELOPMENT LIMITED
blanchardstown, dublin D15 A P2D
EI  D15 AP2D
Manufacturer (Section G)
HARMAC MEDICAL PRODUCTS, INC.
2201 bailey ave.
buffalo NY 14211
Manufacturer Contact
mark wendelken
440 route 22 east
suite 302
bridgewater, NJ 08807
MDR Report Key24378958
Report Number3036773437-2026-00010
Device Sequence Number18767901
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)P308(17)270201
Combination Product (Y/N)Y
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P860003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 02/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCLXUSA
Device Lot NumberP308
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/22/2026
Initial Report FDA Received Date02/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-