| Catalog Number |
175816 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that they had an issue with the bard 16 french foley catheter kit.The balloon did not want to inflate or deflate.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial reporter's phone number: (b)(6).The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation the reported event is addressed within the labeling as it states: visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Note: it is not necessary to pre test the foley catheter balloon.12.Proceed with catheterization in usual manner.When catheter tip has entered bladder, urine will be visible in the drainage tube.Insert catheter two more inches and inflate catheter balloon.13.Inflate catheter balloon using sterile water and fill with recommended volume as referenced on product label.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Corrections: d, g.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that they had an issue with the bard 16 french foley catheter kit.The balloon did not want to inflate or deflate.
|
| |
|
Search Alerts/Recalls
|