| Model Number |
L300 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003, which states that the voltage is too low for projected remaining capacity.It was reported that a request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended.The device still remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003, which states that the voltage is too low for projected remaining capacity.It was reported that a request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended.Additional information was received indicating that this device was explanted and replaced.No additional adverse patient effects were reported, and this device has not been returned.
|
| |
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003, which states that the voltage is too low for projected remaining capacity.It was reported that a request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended.Additional information was received indicating that this device was explanted and replaced.No additional adverse patient effects were reported.This device has been returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
This device was thoroughly inspected and analyzed in our post market quality assurance laboratory.Analysis of the device memory confirmed that a code 1003 alert was recorded.Boston scientific previously issued a field safety notice to notify physicians of the potential for accolade family pacemaker and cardiac resynchronization therapy pacemaker (crt-p) devices to initiate safety mode when high battery impedance is detected.Engineers determined that the code 1003 alert was due to a subsequently released software update that was designed to detect the onset of high battery impedance and prevent the initiation of safety mode in an ambulatory setting.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of message - error code 1003 was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause traced to device design.
|
| |
|