• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACCOLADE? SR; Pacemaker

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACCOLADE? SR; Pacemaker Back to Search Results
Model Number L300
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this device recorded a code 1003, which states that the voltage is too low for projected remaining capacity.It was reported that a request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended.The device still remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Event or Problem Description
It was reported that this device recorded a code 1003, which states that the voltage is too low for projected remaining capacity.It was reported that a request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended.Additional information was received indicating that this device was explanted and replaced.No additional adverse patient effects were reported, and this device has not been returned.
 
Event or Problem Description
It was reported that this device recorded a code 1003, which states that the voltage is too low for projected remaining capacity.It was reported that a request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended.Additional information was received indicating that this device was explanted and replaced.No additional adverse patient effects were reported.This device has been returned for analysis.
 
Additional Manufacturer Narrative
This device was thoroughly inspected and analyzed in our post market quality assurance laboratory.Analysis of the device memory confirmed that a code 1003 alert was recorded.Boston scientific previously issued a field safety notice to notify physicians of the potential for accolade family pacemaker and cardiac resynchronization therapy pacemaker (crt-p) devices to initiate safety mode when high battery impedance is detected.Engineers determined that the code 1003 alert was due to a subsequently released software update that was designed to detect the onset of high battery impedance and prevent the initiation of safety mode in an ambulatory setting.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of message - error code 1003 was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause traced to device design.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE? SR
Common Device Name
Pacemaker
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
+1(651)582-6168
MDR Report Key24381407
Report Number2124215-2026-09184
Device Sequence Number13308267
Product Code LWP
UDI-Device Identifier00802526559150
UDI-Public(01)00802526559150(17)200702
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
N970003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/02/2020
Device Model NumberL300
Device Catalogue NumberL300
Device Lot Number722850
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/17/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2026
Supplement Date Received by Manufacturer03/04/2026
Not provided
Initial Report FDA Received Date02/18/2026
Supplement Report FDA Received Date03/09/2026
08/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/18/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2124215-03/18/26-001-C
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age75 YR
Patient SexMale
-
-