D10 concomitant products: vlocl0336, vlocl0336 v-loc* 180 0 grn 60cm gs21x12 (lot # a4l1463vy); vlocl0336, vlocl0336 v-loc* 180 0 grn 60cm gs21x12 (lot # a4l1463vy); vlocl0336, vlocl0336 v-loc* 180 0 grn 60cm gs21x12 (lot # a4l1463vy); vlocl0336, vlocl0336 v-loc* 180 0 grn 60cm gs21x12 (lot # a4l1463vy).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intraoperatively, a total of 16 sutures broke during use.A new suture from a different lot was tried and used to fix the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.Sixteen vlocl0336 v-loc* 180 0 grn 60cm gs21x12 representative samples were returned for analysis.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that a total of 16 sutures broke during use.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|