While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- solution description (b)(6) 2026, 10:12:22 , (b)(6).Solution description: ptc troubleshot the issue and determined the usb interface needs to be replaced.Added part number 100008290 under warranty.Additional retakes were reported by ptc.Functional (yes/no): no.Customer note (b)(6) 2026 , 11:09:50 , 9b)(6).Troubleshooting: issue reported: not recognized.Did the reported issue require multiple images / multiple exposures to be taken on the patient (yes/no - do not put na):y equipment type and model: schick 3.0 remote serial number: (b)(6) confirm sensor cable colo r:gray acquisition software & version: open dental issue in one or all rooms: yes other sensors work with working remotes in room(s) with issue: yes pc name:op2 schick driver version (programs and features): remote fw/fpga version: n/a sensor fw version: n/a if remote issue: yes 2.0 or 3.0: tried different usb cable/usb port/bypassed usb hubs/extenders: yes 3.0 remote: tried on a 3.0 and 2.0 usb po rt/cable clip connected: yes if sensor issue, replaced elastomer: yes if sensor issue, tried different sensor cable: yes if no response to radiation, confirmed xray head functional: additional troubleshooting steps/notes:.
|