DENTSPLY LLC SCHICK33 S1 SENSOR SHIP KIT/3.0 CABLE 6'; UNIT, X-RAY, INTRAORAL Back to Search Results
Catalog Number 100008627
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- solution description (b)(6) 2026, 10:12:22 , (b)(6).Solution description: ptc troubleshot the issue and determined the usb interface needs to be replaced.Added part number 100008290 under warranty.Additional retakes were reported by ptc.Functional (yes/no): no.Customer note (b)(6) 2026 , 11:09:50 , 9b)(6).Troubleshooting: issue reported: not recognized.Did the reported issue require multiple images / multiple exposures to be taken on the patient (yes/no - do not put na):y equipment type and model: schick 3.0 remote serial number: (b)(6) confirm sensor cable colo r:gray acquisition software & version: open dental issue in one or all rooms: yes other sensors work with working remotes in room(s) with issue: yes pc name:op2 schick driver version (programs and features): remote fw/fpga version: n/a sensor fw version: n/a if remote issue: yes 2.0 or 3.0: tried different usb cable/usb port/bypassed usb hubs/extenders: yes 3.0 remote: tried on a 3.0 and 2.0 usb po rt/cable clip connected: yes if sensor issue, replaced elastomer: yes if sensor issue, tried different sensor cable: yes if no response to radiation, confirmed xray head functional: additional troubleshooting steps/notes:.
 
Event or Problem Description
While the customer was using a part of an intraoral sensor system, schick33 s1 sensor ship kit/3.0 cable 6', they allege experiencing connectivity issues when an x-ray image was taken or an additional image was required.No injury was reported from the alleged event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCHICK33 S1 SENSOR SHIP KIT/3.0 CABLE 6'
Common Device Name
UNIT, X-RAY, INTRAORAL
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key24387692
Report Number2424472-2026-00552
Device Sequence Number10185489
Product Code EAP
UDI-Device IdentifierD764100008627
UDI-PublicD764100008627
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K093453
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number100008627
Was Device Available for Evaluation? No
Distributor Facility Aware Date01/27/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2026
Initial Report FDA Received Date02/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-