| Model Number |
CM3100 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Fa-q325-hf-1 cardiomems duplicate patient profile advisory notice issued by abbott on 20 aug 2025.Capa: 139725 has been initiated to investigate this issue.
|
| |
|
Event or Problem Description
|
|
Abbott r&d internal testing and code review identified an issue where the readings from one cm3100 system session for one patient (patient: (b)(6) were incorrectly reported under a different patient (patient: (b)(6)).No patient injury was reported.The issue is currently under investigation.Mdr: (b)(4) was created for the second patient (patient: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Fa-q325-hf-1 cardiomems duplicate patient profile advisory notice issued by abbott on 20 aug 2025.Capa 139725 has been initiated to investigate this issue.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances were identified that are related to the reported event.
|
| |
|
Search Alerts/Recalls
|