MAUDE Adverse Event Report: GD MEDICAL NS SGL STR TONSIL SPG 1" LARGE; GAUZE/SPONGE, INTERNAL, X-RAY DETECTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GD MEDICAL NS SGL STR TONSIL SPG 1" LARGE; GAUZE/SPONGE, INTERNAL, X-RAY DETECTABLE Back to Search Results
Model Number 30-038NS
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A distributing facility reported through a medwatch, "cotton ball falls off string during use." deroyal industries, inc.Requested return of the device but it was stated it was not available.A supplier corrective action request (scar) has been issued to the supplier, gd medical.This investigation is not complete at this time.No further information is available at this time.We will provide follow up report when additional information becomes available.
 
Event or Problem Description
A distributing facility reported through a medwatch, "cotton ball falls off string during use.".
 
Additional Manufacturer Narrative
A distributing facility reported through a medwatch, "cotton ball falls off string during use." deroyal industries, inc.Requested return of the device but it was stated it was not available.Because the lot number was not provided, the device history record was unable to be reviewed for the affected product.An inventory check was completed on 2 lots of the raw material and no issues were found.A complaint to sales ratio was conducted over the past two years and found to be 0.002%.A supplier corrective action request (scar) has been issued to the supplier, (b)(4).Samples of existing inventory and the work in progress were tested and no duplication of the same issues were identified.Review of process controls of string attachment were also reviewed.Root cause: because the lot number nor sample were returned and no issues were found within the manufacturing process, the root cause could not be determined.Corrective and preventive actions: qa and workshop supervisors were notified of this incident.Reinforced operator training on string attachment process and increased in-process sampling.Periodic verification of attachment strength during production, and ongoing monitoring of compliant trends was also established.Deroyal will also continue to monitor for trends around this item and issue.This investigation is complete at this time.No further information is available at this time.We will provide follow up report if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NS SGL STR TONSIL SPG 1" LARGE
Common Device Name
GAUZE/SPONGE, INTERNAL, X-RAY DETECTABLE
Manufacturer (Section D)
GD MEDICAL
1286 humbracht circle
bartlett IL 60103
Manufacturer (Section G)
DEROYAL INDUSTRIES, INC.
1595 highway 33 south
new tazewell TN 37825
Manufacturer Contact
reagan facemire
200 debusk lane
powell, TN 37849
MDR Report Key24388011
Report Number2320762-2026-00003
Device Sequence Number21404815
Product Code GDY
UDI-Device Identifier00749756162920
UDI-Public00749756162920
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number30-038NS
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/19/2026
Supplement Report FDA Received Date03/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-