| Catalog Number |
02K41-27 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section a patient information: refer to section b5 for complete patient information.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated architect stat troponin-i results generated by the architect i1000sr processing module for a 79-year-old female patient.The lab reference range: 0.00 to 0.04 ng/ml.Sid: (b)(6) (first draw): initial result = 0.55 ng/ml.Repeat result on the same instrument = 0.01 ng/ml.Repeat result on another instrument = 0.001 ng/ml.Sid: (b)(6) (second draw): result = 0.01 ng/ml.Sid: (b)(6) (third draw): result = 0.00 ng/ml.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A complaint investigation was conducted for the architect stat troponin-i, reagent lot 85694un25 to address the customer¿s observation of falsely elevated results.Review of tracking and trending data for the architect stat troponin-i assay lot 85694un25 did identify an increase in complaints, however, the search by list number did not identify an increase in complaint activity for the current complaint issue.Device history review did not identify issues associated with the customer¿s observation.Historical performance for lot 85694un25 was evaluated using worldwide data for the troponin-i assay.The patient median data (divided into 3 groups) and cal f rlu mean were analyzed and compared to an established control limit.This evaluation indicated that all three levels of the patient medians and the cal f rlu mean for lot 85694un25 are within the established limits.Therefore, no unusual reagent lot performance was identified for lot 85694un25.Labeling was reviewed and found to adequately address the customer's issue.Based on our investigation, no systemic issue or deficiency was identified with the architect stat troponin-i, reagent lot 85694un25.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated architect stat troponin-i results generated by the architect i1000sr processing module for a 79-year-old female patient.The lab reference range: 0.00 to 0.04 ng/ml sid (b)(6) (first draw): initial result = 0.55 ng/ml.Repeat result on the same instrument = 0.01 ng/ml.Repeat result on another instrument = 0.001 ng/ml.Sid (b)(6) (second draw): result = 0.01 ng/ml.Sid (b)(6) (third draw): result = 0.00 ng/ml.No impact to patient management was reported.
|
| |
|