• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC 3I T3® NON-PLATFORM SWITCHED TAPERED IMPLANT 4 X 11.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC 3I T3® NON-PLATFORM SWITCHED TAPERED IMPLANT 4 X 11.5MM; DENTAL IMPLANT Back to Search Results
Catalog Number BOST411
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); Osteolysis (2377)
Date of Event 12/16/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided.A2: age and date of birth unknown / not provided.A4: weight unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported that the implant was removed due to peri-implantitis.The patient took antibiotics before the surgery.Symptoms as a result of the event: inflammation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3® NON-PLATFORM SWITCHED TAPERED IMPLANT 4 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key24389900
Report Number0001038806-2026-00866
Device Sequence Number21392470
Product Code DZE
UDI-Device Identifier00844868028932
UDI-Public(01)00844868028932(17)291218(10)2120025007
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K122300
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 02/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberBOST411
Device Lot Number2120025007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient EthnicityNon Hispanic
-
-