MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION LIFE SCOPE TR; BEDSIDE MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION LIFE SCOPE TR; BEDSIDE MONITOR Back to Search Results
Model Number MU-671RA
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that this bedside monitor is having issues with repeatedly picking up a duplicate bed-id with a transport monitor.This appears to be an issue with incorrect use of the wlan transport feature.If the transport is removed from the bsm, both the bsm and transport will attempt to go into wlan transport, at which point the transport takes on the bed-id of the host bsm, once re-docked the bed-id of the transport should be irrelevant as it won't be connecting directly to the network.However, if the transport is taken off the bsm to act standalone, or not in wlan transport, the duplicated bed id from wlan transport does persist and can cause issues.There was no patient injury reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.The following field contains no information (ni), as attempts to obtain the information were made, but not provided: d10 attempt # 1: 02/09/2026 emailed the customer via microsoft outlook for the information in the ni list above: no reply was received.Attempt # 2: 02/11/2026 i called (b)(6) to ask for model and serial numbers of any devices the reported device was connected or related to.A message was left for (b)(6) to call me back with this information.Attempt # 3: 02/13/2026 i called (b)(6) to ask for model and serial numbers of any devices the reported device was connected or related to.A message was left for (b)(6) to call me back with this information.
 
Event or Problem Description
The customer reported that this bedside monitor is having issues with repeatedly picking up a duplicate bed-id with a transport monitor.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFE SCOPE TR
Common Device Name
BEDSIDE MONITOR
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key24389909
Report Number8030229-2026-06205
Device Sequence Number10279066
Product Code MHX
UDI-Device Identifier04931921103562
UDI-Public(01)04931921103562
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K080342
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 05/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMU-671RA
Device Catalogue NumberMU-671RA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2026
Supplement Date Received by Manufacturer05/15/2026
Initial Report FDA Received Date02/19/2026
Supplement Report FDA Received Date05/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-