| Model Number |
GC-322 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a hemorrhoidectomy, at wound closure, the 10 needles were bent when pulled the thread while holding it with the needle holder.The needles detached from the suture thread after only about 3-4 stitches but did not fall into the patient cavity.A new suture was used to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: gc-322, gc-322 caprosyn* 3-0 vio 75cm v20 x36 (lot# d5l2443y) gc-322, gc-322 caprosyn* 3-0 vio 75cm v20 x36 (lot#: d5l2443y) gc-322, gc-322 caprosyn* 3-0 vio 75cm v20 x36 (lot#: d5l2443y) gc-322, gc-322 caprosyn* 3-0 vio 75cm v20 x36 (lot#: d5l2443y) gc-322, gc-322 caprosyn* 3-0 vio 75cm v20 x36 (lot#: d5l2443y) gc-322, gc-322 caprosyn* 3-0 vio 75cm v20 x36 (lot#: d5l2443y) gc-322, gc-322 caprosyn* 3-0 vio 75cm v20 x36 (lot#: d5l2443y) gc-322, gc-322 caprosyn* 3-0 vio 75cm v20 x36 (lot#: d5l2443y) gc-322, gc-322 caprosyn* 3-0 vio 75cm v20 x36 (lot#: d5l2443y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted one needle that was disengaged from its suture.Microscopic inspection showed normal crimp and swage marks, with suture material present in the swage, indicating that the suture broke off at the swage.It was reported that the needles detached from the suture thread.The reported issue was confirmed.The most likely cause could not be established from the information available.It was also reported that the needle was bent.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|