| Model Number |
CDS0706-XTW |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4 functional mitral regurgitation (mr) for a mitraclip procedure.During the procedure, while performing the establish final arm angle (efaa), the clip opened.The clip was then relocked and closed for the second attempt, it failed again.The clip was successfully removed from the patient and replaced with another clip.All steps were performed as per ifu.The final mr was grade 2.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
All available information was investigated, and the reported unintended movement (clip open ¿ efaa) was not confirmed via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.Based on the information provided and the results of the returned device analysis (unable to confirm the reported issue), a cause for the reported unintended movement (clip opened while establishing final arm angle) could not be determined.There is no indication of a product issue with respect to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|