ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0706-XTW
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2026, a patient presented with grade 4 functional mitral regurgitation (mr) for a mitraclip procedure.During the procedure, while performing the establish final arm angle (efaa), the clip opened.The clip was then relocked and closed for the second attempt, it failed again.The clip was successfully removed from the patient and replaced with another clip.All steps were performed as per ifu.The final mr was grade 2.There were no adverse patient effects or clinically significant delays reported.
 
Additional Manufacturer Narrative
All available information was investigated, and the reported unintended movement (clip open ¿ efaa) was not confirmed via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.Based on the information provided and the results of the returned device analysis (unable to confirm the reported issue), a cause for the reported unintended movement (clip opened while establishing final arm angle) could not be determined.There is no indication of a product issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Common Device Name
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL NORTHPOINT REG 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key24390348
Report Number2135147-2026-01027
Device Sequence Number10490459
Product Code NKM
UDI-Device Identifier05415067037381
UDI-Public(01)05415067037381(17)261107(10)51108A1050
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCDS0706-XTW
Device Catalogue NumberCDS0706-XTW
Device Lot Number51108A1050
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2026
Supplement Date Received by Manufacturer02/25/2026
Initial Report FDA Received Date02/19/2026
Supplement Report FDA Received Date03/05/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STEERABLE GUIDE CATHETER.
Patient Age66 YR
Patient SexMale
Patient Weight93 KG
-
-