| Catalog Number |
0475341001 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the shaver was being used to clear the femoral tunnel for a acl revision.It looked as if the shaver had caught on the edge of the tunnel and the inner housing of the shaver had snapped off.
|
| |
|
Event or Problem Description
|
|
It was reported that the shaver was being used to clear the femoral tunnel for a acl revision.It looked as if the shaver had caught on the edge of the tunnel and the inner housing of the shaver had snapped off.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was discarded and not returned for investigation.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.The reported failure mode will be monitored for future reoccurrence.Alleged failure: snapped off.Probable root cause: inadequate window profile.Inadequate welding process.Improper material strength.Inadequate size selected for the application.Material brittleness.Excessive force applied by the user.Incorrect rfid tag.
|
| |
|