MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL DELTAXSFT; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL DELTAXSFT; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number DLX100204
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/31/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer ref# (b)(4).The purpose of this mdr submission is to document the findings of the device investigation.The embolic coil was found to be kinked and stretched, which meets the applicable regulatory reporting criteria.Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section d2b ¿ procode: krd/hcg the device was returned to j&j medtech for further evaluation.A non-sterile deltaxsft10 2mm x 4cm was received contained in the decontamination pouch.Upon receiving the device, visual inspection was performed, and it was noted that the coil was returned protruding from the introducer.Under magnification, the embolic coil was found to be severely stretched on the proximal portion, and the internal blue fiber was exposed.Also, some kinked conditions were noted on ti.However, remained attached to the resistance heating (rh).No other damages were noted in the device.The issue documented a coil that the coil could not be inserted into the microcatheter was confirmed based on the appearance of the returned device.The damage observed in the coil could be the result of the impeded condition experienced during the procedure.Stretching and kinking can occur during procedure handling where force may have been inadvertently applied.According to the risk documentation, friction and difficulty to advance are potential issues that can occur during microcoil placement due to continuous saline flush not being established, which can result in coil damage.Coil stretching and kinking are known potential issues associated with the use of this device.The instructions for use (ifu) provide proper handling instructions for the device to prevent such issues from occurring.There is no indication that the issue reported is a result of a defect inherently related to the device.A manufacturing record evaluation was performed for the finished device 2022553s number, and no internal actions related to the malfunction were identified.As part of j&j medtech quality process, all devices are manufactured, inspected, and released to approved specifications.Therefore, no internal action is required.It should be noted that multiple factors could cause product failure.The instructions for use (ifu) contain the following precautions: if unusual friction is noticed during advancement or retraction of the microcoil system through the introducer, open the rhv main valve, and partially withdraw the distal end of the introducer to expose its tip within the rhv.Tighten the rhv main valve and flush the y-connector of the rhv with sterile saline and verify that fluid exits the slit in the clear portion of the introducer.If unusual friction is still noticed during advancement or retraction of the microcoil system, verify flush lines are open and properly pressurized.Then slowly withdraw the entire microcoil system and examine for damage.Replace it with a new microcoil system.If friction still exists, withdraw and examine the delivery catheter system.If repositioning of the microcoil is necessary, carefully observe the motion of the microcoil in respect to the dpu wire while retracting the microcoil under fluoroscopy.If the microcoil movement is not one-to-one with the dpu wire, or if repositioning is difficult, the microcoil may have become stretched and could possibly break.Gently remove and discard the microcoil system.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an endovascular embolization, two deltafill10 2mm x 8cm coils (product codes: dlf100208, lot numbers 2232554s and 2232541s) and a deltaxsft10 2mm x 4cm coil (product code: dlx100204, lot number: 2022553s) could not be inserted into the echelon10, medtronic microcatheter (mc).Additional event information indicated that another delatafill and deltaxsft coils were successfully used with the microcatheter.There was no need to remove or replace the microcatheter during the procedure.The target vessel was the middle cerebral artery (mca), and specific anatomical characteristics¿including calcification, tortuosity, and vessel diameter¿were not available.Throughout the intervention, there were no observations of physical damage to the devices, and no obstructions within the microcatheter were noted.A continuous flush was maintained for the duration of the procedure.Based on the product analysis of the returned device, the embolic stent was kinked and stretched.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTAXSFT
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC. (FREMONT)
47709 fremont blvd
fremont CA 94538
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key24392872
Report Number3008114965-2026-00310
Device Sequence Number16016852
Product Code KRD
UDI-Device Identifier10886704077329
UDI-Public(01)10886704077329(17)280831(10)2022553S
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDLX100204
Device Lot Number2022553S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2026
Initial Report FDA Received Date02/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ECHELON¿ MICRO CATHETER (MEDTRONIC)
Patient SexUnknown
-
-