|
Product complaint#: (b)(4).Date sent to the fda: 2/19/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.Additional information was requested, the following was obtained: (b)(4).Please provide an answer for each patient-event: it was recently reported that "the doctor did not give other treatment measures for the patient's wound except suture removal." was the suture removed in a routine post-op procedure? or was the suture removed in a reoperation procedure? was reoperation necessary to remove the suture and re-close the wound? was there any medical or surgical intervention performed (product removed; re-operation; re-closure; prescription medication)? if so, please specify.If medication was required, please clarify if it was prescription strength.Please provide the title of external person providing answers to follow-up (e.G.Nurse, surgeon, risk manager).Received information as following from sales rep via phone today.Please refer to the additional event information mentioned on note (b)(4), other information requested is unknown.Was there any medical or surgical intervention performed (product removed; re-operation; re-closure; prescription medication)? if so, please specify.If medication was required, please clarify if it was prescription strength.What is the most current patient status? what is the procedure date (dd/mm/yyyy)? what date did the adverse event occur on (dd/mm/yyyy)? received information as following from sales rep via phone today.Please refer to the event description, other information requested is unknown.The following information was received: per the hospital's feedback, the specific number of events could not be verified, and the involved conditions included: suture breakage easily occurred during intraoperative suturing; the suture could not be withdrawn after postoperative suture removal, some of the suture remained, the wound healing did not meet expectations.No subsequent adverse event reports have been received.After investigation, the patient underwent thyroid tumor resection in the hospital on (b)(6) 2025.The surgeon used vcp397h for tissue suturing during the surgery (the specific suturing tissue level and suturing method were unknown).The intraoperative suturing was smooth, and no device failure or patient injury was reported.After surgery (specific event date unknown), when the patient returned to the hospital for suture removal, the doctor found that the suture could not be withdrawn, and some of the suture remained.The patient had significant scar hyperplasia, wound redness and swelling, and the healing condition did not meet expectations.The doctor did not give other treatment measures for the patient's wound except suture removal.No subsequent adverse event reports were received.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
|
It was reported that a patient underwent a thyroid procedure on an unknown date and suture was used.The suture could not be withdrawn when removing suture after surgery.Some suture remained in body, leading to significant scar hyperplasia.The customer reported that multiple incidents had occurred, but the exact number could not be confirmed.Additional information was requested.
|