| Model Number |
BF-MP190F |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was evaluated.Device evaluation found the device forceps passage had a stuck brush.Cleaning brush passage test failed.Based on investigation results, the reported issue was confirmed.The root cause of the brush stuck in forceps passage inside the channel cannot be conclusively identified.Probable cause based on reasonably known information based on device evaluation was due to blockage caused by kinks/crimped found in the insertion tube of the device.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed during the device evaluation, the bronchofibervideoscope exhibited stuck foreign material inside channel.There was no patient involvement.
|
| |
|