• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 10 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 10 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 10-480
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported "rupture".This record is for the right side.Device remains implanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b5, b6, d6b, h6.
 
Event or Problem Description
Device has been explanted and replaced.
 
Additional Manufacturer Narrative
Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: observed broken device through microscopic inspection assessed as surgical damage and two openings assessed as fold flaw opening.Observed a missing piece of shell through microscopic inspection assessed as inconclusive.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (wear abrasion, non penetrating nick and crease) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Event or Problem Description
Healthcare professional reported "rupture".This record is for the right side.Device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 10 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key24394305
Report Number9617229-2026-03321
Device Sequence Number15943314
Product Code FTR
UDI-Device Identifier10888628001305
UDI-Public10888628001305
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/12/2013
Device Catalogue Number10-480
Device Lot Number1611679
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/29/2026
Supplement Date Received by Manufacturer02/20/2026
04/15/2026
Initial Report FDA Received Date02/19/2026
Supplement Report FDA Received Date03/16/2026
05/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/13/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient SexFemale
Patient Weight58 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-