| Model Number |
SIGTRSB60AXT |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: egiaushort endogia ultra univ sht stap (lot#: unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a distal pancreatectomy, while on pancreatic resection, the firing could not be done at all.The surgeon was able to press the green button.A new cartridge was used with the same handle to resolve the issue.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the reload was pre-fired and the interlock was engaged.Staples were protruding from the proximal end of the staple cartridge channel.The jaw of the reload was open.The sled was slightly advanced.All sutures were intact and the reinforcement material was properly anchored.Functional testing noted that the reload was loaded into a representative handle.The interlock was overridden and the reload was applied to test media.All staples were placed, and test media was transected.The reload interlock was tested and found to function properly.It was reported that the instrument did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if the firing button had been partially compressed and then the open button was pressed after pressing the green firing button.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to completely fire the reload will result in an incomplete cut or incomplete staple formation, which may result in poor hemostasis or leakage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|