The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the objective gluing missing.Based on the result, we concluded that it was caused due to the physical damage applied on the objective gluing.In addition, our technician confirmed that the distal body leak; however, this defect is not the main cause, and/or irrelevant to the alleged complaint.In terms of the objective glue missing, the possibility of dropping into human body could not be denied.Moreover, based on the technical report ""hr-rpt-0974(distal end)"" and/or the risk analysis results, it was evaluated to submit mdr.
|