• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG BACKFLUSH HANDLE DSP; INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED, ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG BACKFLUSH HANDLE DSP; INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED, ACCESSORY Back to Search Results
Catalog Number 337.84
Medical Device Problem Codes Obstruction of Flow (2423); Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
A nurse reported that during vitrectomy surgery an ophthalmic device with tip blockage and no suction.The product was changed to complete surgery.There was no patient harm.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in sections d9, h3, h6 and h11.The returned instrument was received in its inner and outer blister covered with the tyvek foil lid, showing the lot information.The instrument shows no macroscopic signs of damage and no signs of surgery residues.The instrument was visually inspected using a photomicroscope at various magnifications.The visual inspection showed no abnormalities on the soft tip of the product.The instrument was then functionally tested regarding their aspiration/irrigation properties.It could be shown that there is no leakage and no occlusion of the instrument and the aspiration as well as the irrigation works as expected for the instrument.The customer¿s complaint can therefore not be confirmed, as the product met their specifications.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A non-conformance based review of the lot number was performed and did not reveal any potential contributing factors to the reported complaint.A root cause for the customers reported event could not be determined because the returned product met manufacturer¿s specifications.No action was taken as the returned sample met specifications for tests performed in relation to the reported event.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.The complaint data will continue to be monitored for evidence of adverse trending and further action will be taken, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BACKFLUSH HANDLE DSP
Common Device Name
INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED, ACCESSORY
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24399112
Report Number3003398873-2026-00039
Device Sequence Number9730381
Product Code HQE
UDI-Device Identifier07612717071193
UDI-Public(01)07612717071193(17)280430(11)250523(10)17EVF2
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K884043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number337.84
Device Lot Number17EVF2
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/29/2026
Supplement Date Received by Manufacturer03/24/2026
Initial Report FDA Received Date02/20/2026
Supplement Report FDA Received Date04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/23/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-