| Catalog Number |
121732048 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional manufacturer narrative: h3, h6: there is no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.No device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient returned for another component exchange.Patient had extensive head and poly wear a second time.The doctor explanted cup to reposition.Affected side: unknown hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4)
investigation summary no device associated with this report was received for examination.According to the information received, "it was reported that the patient returned for another component exchange.Patient had extensive head and poly wear a second time.Dr explanted cup to reposition.The dr.Would like pinnacle gription cup 121732048 lot 3672827 checked for any known manufacturer defects.Implants were not returned from hospital after requested on date of last surgery.Doi: unknown
dor: (b)(6) 2026
affected side: unknown hip."
product description:- pinn sector w/gription 48mm
product code:- 121732048
lot no:- 3672827
quantity of manufactured:-(b)(4)
date of manufacturing:- 13-jan-2022
expiry date:- 31-dec-2031
ifu reference:- n/a.A manufacturing record evaluation was performed for the finished device product description: pinn sector w/gription 48mm, product code: 121732048, lot number: 3672827 and two non-conformance was identified, however not related to the reported failure mode of the complaint.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot 1) quantity manufactured: 10
2) date of manufacture: 13-jan-2022
3) any anomalies or deviations identified in dhr: two non-conformance was identified, however not related to the reported failure mode of the complaint.4) expiry date: 31-dec-2031
5) ifu reference: n/a.Device history review a manufacturing record evaluation was performed for the finished device product description: pinn sector w/gription 48mm, product code: 121732048, lot number: 3672827 and two non-conformance was identified, however not related to the reported failure mode of the complaint.
|
| |
|