DEPUY ORTHOPAEDICS INC US PINN SECTOR W/GRIPTION 48MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 121732048
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/22/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional manufacturer narrative: h3, h6: there is no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.No device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.
 
Event or Problem Description
It was reported that the patient returned for another component exchange.Patient had extensive head and poly wear a second time.The doctor explanted cup to reposition.Affected side: unknown hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) investigation summary no device associated with this report was received for examination.According to the information received, "it was reported that the patient returned for another component exchange.Patient had extensive head and poly wear a second time.Dr explanted cup to reposition.The dr.Would like pinnacle gription cup 121732048 lot 3672827 checked for any known manufacturer defects.Implants were not returned from hospital after requested on date of last surgery.Doi: unknown dor: (b)(6) 2026 affected side: unknown hip." product description:- pinn sector w/gription 48mm product code:- 121732048 lot no:- 3672827 quantity of manufactured:-(b)(4) date of manufacturing:- 13-jan-2022 expiry date:- 31-dec-2031 ifu reference:- n/a.A manufacturing record evaluation was performed for the finished device product description: pinn sector w/gription 48mm, product code: 121732048, lot number: 3672827 and two non-conformance was identified, however not related to the reported failure mode of the complaint.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot 1) quantity manufactured: 10 2) date of manufacture: 13-jan-2022 3) any anomalies or deviations identified in dhr: two non-conformance was identified, however not related to the reported failure mode of the complaint.4) expiry date: 31-dec-2031 5) ifu reference: n/a.Device history review a manufacturing record evaluation was performed for the finished device product description: pinn sector w/gription 48mm, product code: 121732048, lot number: 3672827 and two non-conformance was identified, however not related to the reported failure mode of the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN SECTOR W/GRIPTION 48MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24399402
Report Number1818910-2026-03326
Device Sequence Number9905444
Product Code KWA
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K093646
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number121732048
Device Lot Number3672827
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/22/2026
Supplement Date Received by Manufacturer07/06/2026
Initial Report FDA Received Date02/20/2026
Supplement Report FDA Received Date07/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-