| Catalog Number |
530.715 |
| Medical Device Problem Codes |
Device Reprocessing Problem (1091); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device the following steps failed: general condition check quick coupling for saw blades during the service evaluation the following failures were identified: functional: leak - liquid visual: contact damage device interaction (2+ devices): cannot engage attachment - coupling conclusion: ¿ failure reported is not confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the battery reciprocator device was leaking a liquid.It was reported in the service order that the device had dirt.It was unknown when the event occurred.There were no reports of delays to a surgical procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2026.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|