As reported by the user facility: microbubbles in arterial side of line.To ensure that there was not a leak coming from the tubing being screwed on to the cvc lumens improperly, facility paused the hd treatment and did a sterile disconnect to aspirate from both lumens of cvc.No air was aspirated, so hd tubing was then reconnected and treatment was resumed.Micro bubbling continued throughout entire treatment.No patient injury reported.
|
|
This report has been identified as b.Braun medical internal report number (b)(4).A review of the discrepancy management system (dsms) database was performed for the reported lot number, and no abnormalities or non-conformances were noted during the in process or final product inspection.One (1) contaminated sample was submitted to the manufacturer for evaluation.Based on the condition of the contaminated sample, no testing could be performed.Although the samples could not be properly tested, based on the complaint description, the reported event is similar to a known issue of air in line during use.The reported issue is due to a defect related to a tubing material change previously implemented.The change in tubing resin can lead to bubbles adhering to the inside of the arterial bloodline, in the segment between the patient connector and the blood pump.These bubbles are not being drawn into the bloodline from the external environment, but rather are the result of degassing of blood on the arterial side of the blood pump due to the high negative pressure within the tubing.With the previous tubing resin, these bubbles remained in the blood and were compressed or dissolved via the positive pressure applied post blood pump.With the new resin, these small bubbles adhere to the inside of the bloodline and may gather together to form larger bubbles.Further investigation of the issue is ongoing, and an action plan will be developed upon completion of the investigation.Urgent medical device correction 2521402-2/17/26-003-c [fda res # 98552] was issued to inform users of the issue and provide information to reduce the risk.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|