| Catalog Number |
00515049500 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during surgery the device would not power on and had no power, and the procedure was completed with another device.There were no consequences or impact to the patient.At investigation it was found that batteries had leaked electrolyte.Diligence is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
This incident has been recorded as an initial/final report under (b)(4).Complaint can be confirmed through the returned product analysis.Functional testing of the returned product found the device would not power on with the original battery pack but did power on and function normally in both modes when connected to a lab battery pack.Visual inspection of the interior of the pack found the batteries had leaked electrolyte inside of the pack.There did not appear to be damage to the wiring of the pack.The device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause has been identified as a manufacturing and/or design issue.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|