• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR INC. ESPRIT¿; SCAFFOLD,INFRAPOPLITEAL,ABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR INC. ESPRIT¿; SCAFFOLD,INFRAPOPLITEAL,ABSORBABLE Back to Search Results
Model Number 1203300-28
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the procedure was to treat a calcified lesion in the anterior tibial artery.The 3.0x28mm esprit btk scaffold was implanted however the scaffold did not appear to be sufficiently apposed to the vessel wall post deployment.No additional treatment was performed.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information provided, a definitive cause for the reported issue could not be determined.Based on the results of the complaint investigation, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESPRIT¿
Common Device Name
SCAFFOLD,INFRAPOPLITEAL,ABSORBABLE
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 2807
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key24401272
Report Number2024168-2026-00713
Device Sequence Number15521234
Product Code NXW
UDI-Device Identifier08717648356995
UDI-Public(01)08717648356995(11)241121(17)261120(10)411216A
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P230036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1203300-28
Device Catalogue Number1203300-28
Device Lot Number411216A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2026
Initial Report FDA Received Date02/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-