| Catalog Number |
383559 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr report replaces mfr report # 9610847-2026-00023 which was filed with the incorrect manufacturing plant selection.Device evaluation: a device history record review was completed by our quality engineer team for provided material number 383559 and lot number 5211932.The review did not reveal any detected abnormalities during the production process that could have contributed to the reported issue, and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect similar issues during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
After attempted placement, a hole was seen in the catheter describe patient /(hcp) / user impact patient was stuck again for iv access other comments: catheter was discarded the hole was in the catheter.The patient had to be stuck twice due to the hole in the catheter.
|
| |
|